FDA Recall Open, Classified

Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System.

Recall: Z-1553-2025 · Initiated March 10, 2025

Recall

Recall Number
Z-1553-2025
Event Number
96435
Firm
Philips North America Llc
FEI Number
3016618143
Product Code
KPR
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
March 10, 2025
Posted
April 11, 2025
Address
222 Jacobs St, Cambridge, MA, 02141-2289

Description

Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System.

Reason

Potential for collimator to fall as a result of incorrect installation.

Action

Consignees received an URGENT MEDICAL DEVICE CORRECTION NOTICE dated 3/7/25. The notice instructs customers of safety precautions to take during scan preparation to ensure devices are fit for use. If consignees notice that the laser from the collimator or crosshairs of the collimator's light field is not able to align with the silkscreen on the front cover of the detector, consignees are to immediately stop use and call their Philips representative. The provided notice is to be posted near affected systems and circulated to all device users. Philips will reach out to consignees to schedule an inspection of devices by field service engineers. Engineers may also replace the collimator, if necessary. Consignees with questions are to call Philips Customer Care Solutions Center at 1-800-722-9377 from Monday through Friday (8:00 AM to 8:00 PM EST).

Distribution

Worldwide - US Nationwide distribution in the states of AK, FL, KS, MS, NH, NY, OR, PA, TX, WI and the countries of Argentina, Australia, Canada, Chile, China, Germany, Hungary, India, Indonesia, Italy, Kenya, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Poland, Romania, Russian Federation, Slovenia, South Africa, Spain, Thailand, United Kingdom, Vietnam.

Quantity

83 units