FDA Recall Terminated

ABX Pentra 400 (version 5.0.8 or lower) The ABX PENTRA 400 and 400C are discrete photometric bench top chemistry analyzers for clinical use. The device is intended to duplicate manual analytical procedures by performing various steps such as pipetting, mixing, heating and measuring color intensity. The device is intended for use in conjunction with certain materials to measure a variety of analytes.

Recall: Z-1553-2016 · Initiated February 19, 2016

Recall

Recall Number
Z-1553-2016
Event Number
73349
Firm
Horiba Instruments Inc
FEI Number
3002698356
Product Code
JJE
Status
Terminated
Root Cause
Software design
Initiated
February 19, 2016
Terminated
August 17, 2016
Address
9755 Research Dr, Irvine, CA, 92618-4626

Description

ABX Pentra 400 (version 5.0.8 or lower) The ABX PENTRA 400 and 400C are discrete photometric bench top chemistry analyzers for clinical use. The device is intended to duplicate manual analytical procedures by performing various steps such as pipetting, mixing, heating and measuring color intensity. The device is intended for use in conjunction with certain materials to measure a variety of analytes.

Reason

Horiba Instruments, Inc. is recalling ABX Pentra 400( version 5.0.8 or lower) and Pentra C400 (version 1.1.2 or lower) because clinical chemistry analyzer malfunctions when the following certain alarms appear on the system.

Action

HORIBA Medical sent an "Urgent Safety Notice" dated February 19, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. A new software version to solve the issue is currently under development; however the firm requests customers use the provided procedure until the new software version is installed on the system, or be extremely vigilant with results of one of the alarms mentioned appears on the system. The letter instructs customers to share the information with laboratory staff and retain the notification as part of laboratory documentation. The letter states it is mandatory for customers to complete and return the enclosed response form within 10 days, in order to verify understanding of it. Any questions or concerns contact the local Horiba Medical representative. For further questions, please call (949) 250-4811.

Distribution

USA (nationwide Distribution) to the states of : CA, NY, AL,OK, MS, MD, NV, MN, TX, KY, WV, MI, NY, MT, IL, FL,WA, ID, MO, WY, NJ, VA, KY, GA, PA, TN, MS, KS, AZ, NE, SC, OR, OH, IA, RI, MI and NC.

Quantity

201