FDA Recall Open, Classified

Atellica CH Revised C Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223.

Recall: Z-1551-2025 · Initiated March 11, 2025

Recall

Recall Number
Z-1551-2025
Event Number
96440
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2432235
Product Code
DCN
Status
Open, Classified
Root Cause
Software Design Change
Initiated
March 11, 2025
Posted
April 10, 2025
Address
511 Benedict Ave, Tarrytown, NY, 10591-5005

Description

Atellica CH Revised C Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223.

Reason

Incorrect software flagging may occur for the assay that may potentially lead to an erroneous result.

Action

On March 11, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Customer Actions " Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. " For both Atellica CH and Atellica CI analyzers, perform the instructions in Appendix B to temporarily reduce the measuring interval. o Until the measuring interval is restored, track additional reagent consumption as a result of these actions to report to Siemens Healthineers for future reimbursement/credit. " Additionally, for Atellica CH Analyzers, perform the instructions in Appendix C to remove rules for flagging of No Calculation results and to install Atellica Solution Software version 1.29.0 or higher. " Complete and return the Field Correction Effectiveness Check form attached to this letter within 30 days. " Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution A follow-up communication will be provided when Customer Actions are no longer required. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.

Distribution

Domestic: Nationwide Distribution. International: Australia, Austria, Bahamas, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Republic, Denmark,¿Estonia, Finland, France, Germany, Greece,¿Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Latvia, Lithuania, Malaysia, Maldives, Netherlands, New Zealand, Norway, Oman,¿Pakistan, Poland, Portugal, Qatar, Romania, Saudi¿Arabia, Serbia, Singapore, Slovakia, Slovenia, South¿Africa, Spain, Sweden, Switzerland, Taiwan, U.A.E., Uganda, United¿Kingdom, Uzbekistan, Vatikancity.¿ ¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿ ¿¿¿¿¿¿¿ ¿ ¿¿¿¿ ¿¿¿¿¿¿ ¿¿¿¿¿¿¿¿¿¿ ¿¿¿¿¿¿¿¿ ¿¿¿¿¿ ¿ ¿¿¿¿¿¿