FDA Recall Terminated

AirLife Resuscitation, Adult, with mask, oxygen reservoir bag, manometer Product Usage: The AirLife Resuscitation devices are manual resuscitation devices that are used to provide positive pressure ventilation to patients who are not breathing or cannot adequately breathe on their own.

Recall: Z-1551-2018 · Initiated April 27, 2017

Recall

Recall Number
Z-1551-2018
Event Number
78412
Firm
Vyaire Medical
FEI Number
2050001
Product Code
OEV
Status
Terminated
Root Cause
Component design/selection
Initiated
April 27, 2017
Posted
May 10, 2018
Terminated
May 9, 2019
Address
75 N Fairway Dr, Vernon Hills, IL, 60061-1845

Description

AirLife Resuscitation, Adult, with mask, oxygen reservoir bag, manometer Product Usage: The AirLife Resuscitation devices are manual resuscitation devices that are used to provide positive pressure ventilation to patients who are not breathing or cannot adequately breathe on their own.

Reason

The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.

Action

The firm, Vyaire Medical, sent an "URGENT RECALL NOTICE" letter on 04/27/2017 and again on 06/08/2017 for the expansion of the recall to its Customers. The letter described the product, problem and actions to be taken. The customers were instructed to inspect current inventory on-hand and immediately perform a 1OO% physical inventory of the product to identify and be removed; complete and return the Customer Response Form by Fax to: (312) 949.0972 or email to:[email protected].; destroy all affected product in accordance with your facility's destruction protocol; and for replacements, contact your distributor Armstrong Medical direct at (800)323.4220 x129. For any additional questions and support concerning this voluntary recall, please contact Customer Advocacy Analyst at (224) 706.6829 or email: [email protected].

Distribution

Worldwide Distribution - US (Nationwide) and in the countries of Canada, Puerto Rico, United Arab Emirates, Philippines, and Europe.

Quantity

30 units