FDA Recall Terminated

Peloris Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly, 3149 VIC, Australia; catalog number 26.0001 and 26.0005

Recall: Z-1551-2010 · Initiated February 9, 2010

Recall

Recall Number
Z-1551-2010
Event Number
54566
Firm
Leica Microsystems, Inc.
FEI Number
3000210133
Product Code
IEO
Status
Terminated
Root Cause
Other
Initiated
February 9, 2010
Posted
May 7, 2010
Terminated
June 30, 2011
Address
2345 Waukegan Rd, Bannockburn, IL, 60015-1515

Description

Peloris Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly, 3149 VIC, Australia; catalog number 26.0001 and 26.0005

Reason

Confusion during manually changing reagents, general users running unvalidated protocols, and wax valve leakage have contributed to tissue damage during processing, leading to rebiopsies.

Action

Leica Microsystems sent Medical Device Correction letters dated 2/9/10 to the affected customers via first class mail on the same date, informing them of the potential for tissue damage during processing and outlining the software upgrade being made to minimize the risk of tissue damage. Customers were informed that their local Leica Microsystems representative will contact them to arrange the software upgrade and to provide training on the use of the new software. The letters also provided the necessary steps required to avoid tissue damage during processing while they are awaiting the software upgrade. The accounts were asked to complete the attached Acknowledgement Form, listing the serial numbers of the units at their facility and fax it back to Leica at 1-847-607-3147. Additional questions should be addressed to the firm's office of Regulatory Affairs and Quality Assurance.

Distribution

Worldwide distribution -- U.S. (nationwide), Canada, Australia, Europe and Asia.

Quantity

574 units