BD Pyxis MedBank CUBIE Replenishment Station Product / Material ID BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 155141-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 155140-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 155142-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 155143-01 BD PYXIS MEDBANK TWR MN CR-14HH-P / 155144-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 155145-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 155156-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 155157-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P / 155151-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 155152-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 155153-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 155154-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 155155-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 155146-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 155147-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 155148-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 155149-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 155150-01 BD PYXIS MEDBANK MINI CR-2HH-1FH-P / 169-114 BD PYXIS MEDBANK MINI CR-4HH-P / 169-115 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 169-82 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 169-83 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 169-84 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 169-85 BD PYXIS MEDBANK TWR MN CR-14HH-P / 169-86 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 169-87 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 169-88 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 169-89 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 169-90 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 169-91 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 169-92 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P /169-93 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 169-94 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 169-95 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 169-96 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 169-97 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 169-98 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 169-99 BD PYXIS MEDBANK MINI CR-2HH-1FH-P / 138913-01 BD PYXIS MEDBANK MINI CR-4HH-P / 138911-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FH-P / 139001-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FM-P / 139002-01 BD PYXIS MEDBANK MN 200 CR-4HH-P / 139000-01 BD PYXIS MEDBANK MN 500 CR-2HH-1FM-2DD-P / 139020-01 BD PYXIS MEDBANK MN 500 CR-2HH-5FM-P / 139018-01 BD PYXIS MEDBANK MN 500 CR-4HH-1FH-3FM-P / 139019-01 BD PYXIS MEDBANK MN 500 CR-4HH-4FM-P / 139021-01 BD PYXIS MEDBANK MN 500 CR-6HH-3FM-P / 139022-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 138973-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 138975-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 139073-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 138976-01 BD PYXIS MEDBANK TWR MN CR-14HH-P / 138977-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 138951-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 139072-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 139085-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 138968-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 138964-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 138966-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P /138967-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 138969-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 138965-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 138971-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 138972-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 138974-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 138970-01 BD Pyxis QFill Replenishment Station / 138904-01 KIT BD PYXIS QFILL REPLENISHMENT STATION / 155197-01 MEDBANK UPOS211 AIO WIN10 QFILL 1.5 / 155215-05 OPT,MEDBANK,HD,WIN10,QFILL,CR / 131944-04 The BD Pyxis" MedBank" CUBIE" Replenishment Station is an Automated Dispensing Cabinet (ADC) that is intended to be used by a qualified and authorized user to replenish medications and supplies to a BD Pyxis" MedBank" cabinet.
Recall
- Recall Number
- Z-1550-2025
- Event Number
- 96310
- Firm
- CareFusion 303, Inc.
- FEI Number
- 2016493
- Product Code
- BRY
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- March 11, 2025
- Posted
- April 9, 2025
- Address
- 10020 Pacific Mesa Blvd, San Diego, CA, 92121-4386
Description
BD Pyxis MedBank CUBIE Replenishment Station Product / Material ID BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 155141-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 155140-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 155142-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 155143-01 BD PYXIS MEDBANK TWR MN CR-14HH-P / 155144-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 155145-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 155156-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 155157-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P / 155151-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 155152-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 155153-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 155154-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 155155-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 155146-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 155147-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 155148-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 155149-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 155150-01 BD PYXIS MEDBANK MINI CR-2HH-1FH-P / 169-114 BD PYXIS MEDBANK MINI CR-4HH-P / 169-115 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 169-82 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 169-83 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 169-84 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 169-85 BD PYXIS MEDBANK TWR MN CR-14HH-P / 169-86 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 169-87 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 169-88 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 169-89 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 169-90 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 169-91 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 169-92 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P /169-93 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 169-94 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 169-95 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 169-96 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 169-97 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 169-98 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 169-99 BD PYXIS MEDBANK MINI CR-2HH-1FH-P / 138913-01 BD PYXIS MEDBANK MINI CR-4HH-P / 138911-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FH-P / 139001-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FM-P / 139002-01 BD PYXIS MEDBANK MN 200 CR-4HH-P / 139000-01 BD PYXIS MEDBANK MN 500 CR-2HH-1FM-2DD-P / 139020-01 BD PYXIS MEDBANK MN 500 CR-2HH-5FM-P / 139018-01 BD PYXIS MEDBANK MN 500 CR-4HH-1FH-3FM-P / 139019-01 BD PYXIS MEDBANK MN 500 CR-4HH-4FM-P / 139021-01 BD PYXIS MEDBANK MN 500 CR-6HH-3FM-P / 139022-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 138973-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 138975-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 139073-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 138976-01 BD PYXIS MEDBANK TWR MN CR-14HH-P / 138977-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 138951-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 139072-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 139085-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 138968-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 138964-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 138966-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P /138967-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 138969-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 138965-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 138971-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 138972-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 138974-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 138970-01 BD Pyxis QFill Replenishment Station / 138904-01 KIT BD PYXIS QFILL REPLENISHMENT STATION / 155197-01 MEDBANK UPOS211 AIO WIN10 QFILL 1.5 / 155215-05 OPT,MEDBANK,HD,WIN10,QFILL,CR / 131944-04 The BD Pyxis" MedBank" CUBIE" Replenishment Station is an Automated Dispensing Cabinet (ADC) that is intended to be used by a qualified and authorized user to replenish medications and supplies to a BD Pyxis" MedBank" cabinet.
Due to a software issue, when attempting to restock the automated dispensing cabinet, user have received the banner message "Stocking window has passed, or item has been removed from purchase order. Please contact Pharmacy for assistance". When this occurs, the medication cannot be restocked and can potentially lead to delays in accessing medication for patients.
On 03/11/2025, firm sent via FedEx 2-day mail and email an "URGENT MEDICAL DEVICE CORRECTION" Letter to customer informing them of 3 failure scenarios associated with: -BD Pyxis MedBank CUBIE Replenishment (CR) Station -BD Pyxis MedBank Mini -BD Pyxis MedBank Tower -BD Pyxis MedFlex -BD Pyxis MedBank MedPass Where failure of these systems could result in inaccuracies or discrepancies in inventory management of medications. Customer are instructed to: 1.Best practice is to verify the total dose being removed is accurate against the provider s order, and if available utilize barcode medication administration as an additional check to ensure the correct dose is retrieved prior to administration.1 2.Customers can inventory the CUBIE" pocket via a cycle count or blind count to correct the inventory when discrepancies occur. Customers can refer to the BD Pyxis MedBank Cabinet User Guide - 10000414987 Ver. 01, Version v2.4 for instructions. Refer to chapter six (6), page 103. 3.Customers should ensure backup procedures (which may include but are not limited to keys, emergency medication supplies, alternate devices in proximity, etc) are in place and staff have all been properly educated on them. 4.Please share this letter with any facilities within your network that you may have distributed affected devices to. 5.Please post this Recall Letter on or near affected products as a reminder of the recall and the steps/instructions that are needed to be followed. 6.Complete the attached Response Form and return per the instructions noted on the form even if you do not you have any of the recalled products so that BD may acknowledge your receipt of this recall notification. For questions or assistance: email [email protected] or Phone 1-866-930-9251 24 hour, 7 days a week
U.S.: AL,AZ, CA, CT, FL, GA, IA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, PA, TN, TX, UT, VA, WA, WI and WV. O.U.S.: None
404 systems