Roche Elecsys Folate II for use on Elecsys 210 Analyzers; Catalog No. 3253678 (Roche Material No. 03253678122.
Recall: Z-1550-06
·
Initiated September 15, 2006
Recall
- Recall Number
- Z-1550-06
- Event Number
- 36157
- Firm
- Roche Diagnostics Corp.
- FEI Number
- 1823260
- Product Code
- CGN
- Status
- Terminated
- Root Cause
- Other
- Initiated
- September 15, 2006
- Posted
- September 29, 2006
- Terminated
- April 30, 2008
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025
Description
Roche Elecsys Folate II for use on Elecsys 210 Analyzers; Catalog No. 3253678 (Roche Material No. 03253678122.
Reason
Gel-like particle formation; may result in a clot settling in the analyzer measuring cell, which will create a low signal and may cause incorrectly high or low QC and/or patient results depending on the assay method.
Action
Consignees were informed via letter dated 9/15/06 to perform Folate assays in batch mode and to run controls and liquid flow cleaning upon completion of each batch run. Distributors were requested to notify their customers.
Distribution
Nationwide.
Quantity
5688 units