Gendex VixWin Platinum Imaging Software, Versions 1.0, 1.1 and 1.2; Gendex Dental Systems, 901 West Oakton Street, Des Plaines, IL 60018. Intended for use of the software is to control capture, display, treatments, analysis and saving of x-ray digital images.
Recall
- Recall Number
- Z-1548-2009
- Event Number
- 51485
- Firm
- Gendex Dental Systems
- FEI Number
- 3004115000
- Product Code
- MUH
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- March 6, 2009
- Posted
- July 17, 2009
- Terminated
- January 7, 2011
- Address
- 901 W Oakton St, Des Plaines, IL, 60018-1843
Description
Gendex VixWin Platinum Imaging Software, Versions 1.0, 1.1 and 1.2; Gendex Dental Systems, 901 West Oakton Street, Des Plaines, IL 60018. Intended for use of the software is to control capture, display, treatments, analysis and saving of x-ray digital images.
There is a compatibility issue when the VixWin Platinum versions 1.0/1.1/1.2 software is used in conjunction with Easy Dental, PracticeWorks, EagleSoft and SoftDent practice management software, resulting in patient files that may contain images that are mixed between patients.
Gendex issued an "Urgent Medical Device Report" dated March 6, 2009 to the direct account dealers of the affected device via first class mail. The noticed described the issue and provided a software update directed to fixing the problem. In addition, customers were asked to complete and return an Acknowledgement Form by fax to 1-847-718-0601. Further questions should be directed to the Gendex Technical Support team at 1-888-275-5286.
Nationwide Distribution.
1,256 units