Olympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units to cut tissue using high-frequency current within the upper digestive tract Model: KD-640L

Recall: Z-1547-2024 · Initiated March 1, 2024

Recall

Recall Number
Z-1547-2024
Event Number
94173
Firm
Olympus Corporation of the Americas
FEI Number
2429304
Product Code
KNS
Status
Open, Classified
Root Cause
Use error
Initiated
March 1, 2024
Posted
September 12, 2024
Address
3500 Corporate Pkwy, Center Valley, PA, 18034-8229

Description

Olympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units to cut tissue using high-frequency current within the upper digestive tract Model: KD-640L

Reason

Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions for Use, which details critical specifications regarding electrosurgical unit compatibility and output due to reports of serious injury

Action

Olympus Corporation of the Americas issued Medical Device Safety Reminder letter on 3/1/24 via UPS. Letter states reason for recall, health risk and action to take: 1. Follow your facility's procedures for communication and handling of Field Safety Notices. Ensure all personnel, including clinical staff, are informed of the contents of this letter and the Instructions for Use. You may add a copy of this letter with your IFU. 2. Access the Olympus recall portal to indicate that you have received this notification. a. Go to https://olympusamerica.com/recall. b. Enter the recall number: 0443 c. Complete the form as instructed. 3. If you have distributed these devices outside your facility, please provide a copy of this letter to those facilities immediately. Olympus requests you to report any complaints, including any injuries associated with Triangle Tip Electrosurgical knife tip breakage, the Olympus Technical Assistance Center at 1-800-848-9024, option 1. If you require additional information, please do not hesitate to contact me directly at 647-999-3203 or [email protected].

Distribution

Nationwide

Quantity

4507 units