Alcohol Control sold by Pointe Scientific; Catalog # 7-A7504-CTL.
Recall
- Recall Number
- Z-1546-06
- Event Number
- 35499
- Firm
- Pointe Scientific, Inc.
- FEI Number
- 1827821
- Product Code
- JIX
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 9, 2006
- Posted
- September 30, 2006
- Terminated
- November 20, 2007
- Address
- 5449 Research Dr, Canton, MI, 48188-2261
Description
Alcohol Control sold by Pointe Scientific; Catalog # 7-A7504-CTL.
Open vial stability-Potential microbial growth in the alcohol standard and possible unexpected QC changes in the ammonia portion of the control due to the lack of validation for long-term open vial stability.
Consignees were notified via recall letter dated 6/9/06 and a second letter dated 6/21/06 extending the recall to additional products/lots, to cease using the products, conduct a recall to the user level, destroy products on hand and to notify the firm via the response form of the amount destroyed for replacement purposes.
Worldwide distribution ---- including USA and countries of Canada, Ecuador, Germany, Greece, India, Mexico, Peru, Poland, Portugal, Trinidad, United Arab Emirates and United Kingdom.
44 units