FDA Recall Open, Classified

LINK SLED Knee System POREX ENDO SLED KNEE FEM COMP, LARGE Item Number: 15-2220/60

Recall: Z-1542-2023 · Initiated March 21, 2023

Recall

Recall Number
Z-1542-2023
Event Number
92088
Firm
Linkbio Corp.
FEI Number
3006721341
Product Code
KRO
Status
Open, Classified
Root Cause
Labeling design
Initiated
March 21, 2023
Posted
May 4, 2023
Address
101 Round Hill Dr, Ste 7, Rockaway, NJ, 07866-1214

Description

LINK SLED Knee System POREX ENDO SLED KNEE FEM COMP, LARGE Item Number: 15-2220/60

Reason

Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected

Action

LinkBio issued Urgent Medical Device Correction letter on March 21, 2023 to US Distributors. Letter states reason for recall, health risk and action to take: Please take note of the amendment / reinforcement of the Surgical Technique and participate in future offering of product training on the updated technique. Two warnings have been added. No product return is required. Please respond using the attached Distributor Reply form to indicate your understanding of this Device Correction and of the need to follow the updated surgical technique. Attendance in future product training on the LINK SLED is required for continued use of the system. No action is required for devices which have already been implanted; continue with regular follow-up activities for patients who have received the LINK SLED knee system. All product lots of the above item numbers still within the labeled expiration date are in scope for this device correction. Waldemar Link GmbH & Co. KG is updating the surgical technique to incorporate the new warnings. The company will also be providing product training, through the LinkBio Medical Education team. Please contact LinkBio at 973-625-1333 or [email protected] for any questions.

Distribution

NY, NV Foreign: Bulgaria, France, Italy, Norway, Poland, Sweden, Switzerland, Slovakia, Spain, United Kingdom, Germany, Austria, Argentina, Australia, China, India, Iraq, Jordan, Colombia, Mexico, Saudi Arabia, Ukraine, Vietnam, and Belarus

Quantity

5 units