Revogene RNA Software Upgrade Kit Revogene RNA Software Upgrade USBs Catalog Number: 610240 IVD used to install the Revogene System Software version 4.2.0, import the Revogene Instrument Control Protocol file for RNA-based assay, and install the Assay Definition File (ADF) file
Recall
- Recall Number
- Z-1541-2021
- Event Number
- 87556
- Firm
- Meridian Bioscience Inc
- FEI Number
- 1524213
- Product Code
- QJR
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- April 8, 2021
- Terminated
- October 26, 2021
- Address
- 3471 River Hills Dr, Cincinnati, OH, 45244-3023
Description
Revogene RNA Software Upgrade Kit Revogene RNA Software Upgrade USBs Catalog Number: 610240 IVD used to install the Revogene System Software version 4.2.0, import the Revogene Instrument Control Protocol file for RNA-based assay, and install the Assay Definition File (ADF) file
Emergency Use Application (EUA) application withdrawn as false negative results reported
Meridan issued Urgent Medical Device Recall on 4/08/21 to the user level via phone and electronic mailing . Letter states reason for recall, health risk and action to take: Revogene SARS-CoV-2, Catalog 410700: Laboratorys should consider evaluating patient test results per laboratory policies and procedures. Review current inventory and segregate any remaining stock. Contact Meridian Technical Support 800.343.3858 to obtain a FedEx label to return any remaining unexpired inventory to Meridian Bioscience Inc. Any expired inventory should be destroyed in accordance with facility protocols. Complete and return the Customer Notification Form. This will indicate receipt of this field correction notice. Revogene RNA Software Upgrade Kit, Catalog 610240: Review current inventory and segregate any remaining stock of the RNA Software kit, including the USB. Contact Meridian Technical Support 800.343.3858 to obtain a FedEx label to return remaining unexpired inventory to Meridian Bioscience, Inc. Any expired inventory should be destroyed in accordance with facility protocols. Complete and return the Customer Notification Form. This will indicate receipt of this field correction notice. Actions to Be Taken By Meridian: Meridian intends to conduct additional validation and bridging studies with the intention of re-submitting an EUA application for its SARS-CoV-2 test on the Revogene platform as soon as possible thereafter. Contact Information: If you have any questions, please call Meridian Bioscience Technical Services at 1-800-343-3858, or email at [email protected].
Nationwide
124 Revogene RNA Software Upgrade Kits