FDA Recall Terminated

Dimension Vista(R) Total Prostate Specific Antigen Flex(R) reagent cartridge Product Usage: The TPSA method is an in vitro diagnostic test for the quantitative measurement of total prostate specific antigen (PSA) in human serum and plasma on the Dimension Vista(R) System

Recall: Z-1540-2012 · Initiated April 2, 2012

Recall

Recall Number
Z-1540-2012
Event Number
61629
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
MTF
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
April 2, 2012
Posted
May 11, 2012
Terminated
April 2, 2013
Address
500 GBC Drive PO BOX 6101, Mailstop 514, Newark, DE, 19714-6101

Description

Dimension Vista(R) Total Prostate Specific Antigen Flex(R) reagent cartridge Product Usage: The TPSA method is an in vitro diagnostic test for the quantitative measurement of total prostate specific antigen (PSA) in human serum and plasma on the Dimension Vista(R) System

Reason

Siemens has confirmed a report of leaking Flex(R) reagent cartridge lot 11300BB from Dimension Vista(R) TPSA; when a leaking Flex(R) is used, patient test results may be falsely depressed.

Action

Siemens sent an Urgent Field Safety Notification letter dated April 14, 2012 to all Dimension Vista(R) System customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately discontinue use and discard any remaining inventory of the affected product. Customers were asked to complete the attached form and fax it to 302-631-8467 to indicate that you have received the notification. For technical questions or concerns, contact the Siemens Technical Solutions Center at 800-441-9250.

Distribution

USA Nationwide Distribution including the states of: AL, AZ, CA, CO, CT, FL IL, LA, MD, MI, MN, MS, MT, NC, NH, NJ, NV, NY, OH, PA, SD, TN, TX, UT, WI ,WV including Puerto Rico.

Quantity

1,103 cartons