FDA Recall Terminated

AEM Disposable Electrode, Spatula, 35cm, REF ES0301, Qty: 10 each Sterile R, Manufactured for: Encision Inc., Boulder, CO 80301, made in Taiwan. Electrosurgical accessories intended, by use of monopolar high-frequency electrical current from compatible electrosurgical generators, for ablation, removal, resection and coagulation of soft tissue where associated hemostasis is required in open, endoscopic and laparoscopic surgical procedures.

Recall: Z-1539-2010 · Initiated April 5, 2010

Recall

Recall Number
Z-1539-2010
Event Number
55286
Firm
Encision, Inc.
FEI Number
1000117361
Product Code
GEI
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
April 5, 2010
Posted
May 6, 2010
Terminated
May 21, 2010
Address
6797 Winchester Cir, Boulder, CO, 80301-3513

Description

AEM Disposable Electrode, Spatula, 35cm, REF ES0301, Qty: 10 each Sterile R, Manufactured for: Encision Inc., Boulder, CO 80301, made in Taiwan. Electrosurgical accessories intended, by use of monopolar high-frequency electrical current from compatible electrosurgical generators, for ablation, removal, resection and coagulation of soft tissue where associated hemostasis is required in open, endoscopic and laparoscopic surgical procedures.

Reason

Electrosurgical electrode tip insulation may crack and break off during use.

Action

A "Confirmation of Product Recall" letter dated April 5, 2010, was faxed by the firm (Encision, Inc.) to all customers. The letter describes the product, problem and action to be taken by customers. The customers were instructed to either return or destroy the affected electrodes and to contact any customers or sales representatives to whom they distributed the product and confirm they have been removed from use and/or they will pick them up for return. The customers is to complete form and fax to the attention of RA/QA at 1-303-444-2693 or return with the items. The number to call to arrange return and/or if you have any questions is 1-800-998-0986.

Distribution

Worldwide distribution: USA including states of CA, CT, IA, IN, KS, KY, MI, OK, TX, WV, and countries of New Zealand and The Netherlands.

Quantity

260 units