XSYSTEMS AxSYM Multiconstituent Controls for Abused Drug Assays, list 9687-12; Abbott laboratories, Abbott Park, IL 60064 USA
Recall
- Recall Number
- Z-1539-05
- Event Number
- 32540
- Firm
- Abbott Laboratories MPG
- FEI Number
- 1415939
- Product Code
- DIF
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 27, 2005
- Posted
- September 14, 2005
- Terminated
- January 18, 2006
- Address
- 100 Abbott Park Rd, Abbott Park, IL, 60064-3502
Description
XSYSTEMS AxSYM Multiconstituent Controls for Abused Drug Assays, list 9687-12; Abbott laboratories, Abbott Park, IL 60064 USA
The XSYSTEMS Multiconstituent Controls accessory insert (commodity number 34-3067/R6) does not contain the correct control ranges for the AxSYM Opiates assay.
On 6/27/05, Abbott sent the customer letter via FedEx to all XSYSTEMS Multiconstituent Controls customers that received List Number 9687-12 lots 17270M100 and/or 18177M300 and/or AxSYM Opiates reagents (List Number 3B25-20) since June 01, 2004. The letters explained that the multiconstituent control ranges for morphine were not updated in the MCC Accesory insert at the same time they were updated in the AxSYM Opiates reagent package insert in 2004. The accounts were requested to review their current control ranges for the AxSYM Opiates assay, and verify that they are using either the control ranges in the AxSYM Opiates reagent package insert or control ranges established by their facility. They were also instructed to follow their facility''s normal policies and procedures for communicating with health care providers served by their laboratory. The accounts were requested to complete and return the enclosed reply sheet acknowledging the receipt of the letter and the dissemination of the letter to the laboratory manager/supervisor responsible for the assays.
Nationwide, including Puerto Rico, and internationally to Argentina, Australia, Barbados, Bermuda, Brazil, Canada, Cayman Island, Chile, China/Hong Kong, Columbia, Costa Rica, Dominican Republic, Ecuador, Guatemala, Germany, Honduras, Jamaica, Korea, Mexico, New Zealand, Panama, Peru, Singapore, St. Vincent, Thailand, Trinidad and Taiwan
4151 kits