FDA Recall Terminated

Dimension Vista CHEM 3 calibrator is an in vitro diagnostic product for the calibration of ammonia (AMON), Cardon Dioxide (CO2) and Ethyl Alcohol (ETOH) on the Dimension Vista System

Recall: Z-1535-2010 · Initiated October 23, 2009

Recall

Recall Number
Z-1535-2010
Event Number
53836
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
JIX
Status
Terminated
Root Cause
Other
Initiated
October 23, 2009
Posted
May 4, 2010
Terminated
June 22, 2010
Address
500 GBC Drive PO BOX 6101, Mailstop 514, Newark, DE, 19714-6101

Description

Dimension Vista CHEM 3 calibrator is an in vitro diagnostic product for the calibration of ammonia (AMON), Cardon Dioxide (CO2) and Ethyl Alcohol (ETOH) on the Dimension Vista System

Reason

patient results are depressed

Action

An "Urgent Field Safety Notice" dated October 2009 was issued by Siemens to all customers. The notice described the product, problem and actions to be taken by the customers. The customers were instructed to recalibrate the AMON method (AMON Flex reagent cartridge, K3019) with Vista Chem 3 Claibrator lot 9ED058 using the revised bottle values provided in the notice and to please consult with their laboratory director regarding re-evaluation of previous results. The customers should forward this notification to anyone to whom they may have distributed this product. If you have any technical questions regarding this information, please contact Siemens Technical Solutions Center at 1-800-441-9250.

Distribution

Worldwide distribution: USA and Australia, Canada, Japan, New Zealand, South Korea, and Switzerland.

Quantity

1753 cartons