FDA Recall Terminated

TransFx External Fixation System Drill Bit, Quick Connect, 2.0 mm diameter, 125 mm length, nonsterile, Zimmer, Warsaw, IN; REF 4450-54. TransFx drill bits can be found in the following kits: TransFx Intermediate External Fixation System, REF 4450-05-10; TransFx Large External Fixation System, REF 4450-010-10 and TransFx Small External Fixation System, REF 4450-00050-15. For use in pre-drilling bone to accept a fixation pin, which are part of the external fixation structure that the surgeon assembles intra-operatively to stabilize bone fractures from the extremities.

Recall: Z-1532-2009 · Initiated May 1, 2009

Recall

Recall Number
Z-1532-2009
Event Number
51924
Firm
Zimmer Inc.
FEI Number
1000220733
Product Code
LXT
Status
Terminated
Root Cause
Process control
Initiated
May 1, 2009
Posted
June 30, 2009
Terminated
November 3, 2010
Address
345 E Main St, Warsaw, IN, 46580-2746

Description

TransFx External Fixation System Drill Bit, Quick Connect, 2.0 mm diameter, 125 mm length, nonsterile, Zimmer, Warsaw, IN; REF 4450-54. TransFx drill bits can be found in the following kits: TransFx Intermediate External Fixation System, REF 4450-05-10; TransFx Large External Fixation System, REF 4450-010-10 and TransFx Small External Fixation System, REF 4450-00050-15. For use in pre-drilling bone to accept a fixation pin, which are part of the external fixation structure that the surgeon assembles intra-operatively to stabilize bone fractures from the extremities.

Reason

Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation.

Action

Distributors and consignees were notified by an Urgent: Device Recall letter dated 5/1/09. Consignees were advised to stop using the device and quarantine it immediately. Distributors were instructed to carry out a physical count of all affected product and record the data on the Inventory Return Certification Form; fax a copy of the completed form to Zimmer, Inc.; and return recalled product along with the Inventory Return Certification Form to the firm.

Distribution

Worldwide Distribution -- USA, Australia, Canada, El Salvador, England, Germany, Latin America, and Singapore.

Quantity

1206