VIA Medical Collection Bag; The collection bag is an accessory waste bag used in conjunction with the sterile LVM blood gas sensor with the LVM Monitor. The collection bag is packaged in a tyvek pouch which is included in either replacement kit (777-3103) or LVM sensor kit (777-3102).
Recall
- Recall Number
- Z-1531-2013
- Event Number
- 65341
- Firm
- International Biomedical, Ltd.
- FEI Number
- 1000058335
- Product Code
- LZF
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- May 24, 2013
- Posted
- June 11, 2013
- Terminated
- July 29, 2013
- Address
- 8508 Cross Park Dr, Austin, TX, 78754-4532
Description
VIA Medical Collection Bag; The collection bag is an accessory waste bag used in conjunction with the sterile LVM blood gas sensor with the LVM Monitor. The collection bag is packaged in a tyvek pouch which is included in either replacement kit (777-3103) or LVM sensor kit (777-3102).
International Biomedical has received reports of interference between the collection bag and the extension set connectors. Evaluation of the affected components indicates that the collection bag connector can become occluded due to an over-tightening of the luer lock by the end user. To alleviate the occurrence of this issue in the fixture, the connector on the collection bag has been improved to
International Biomedical decided to recall and sent "Product Bulletin" notification letters to their consignees on 05/24/2013 or 05/30/2013. The letter described the product issue and provided customers with recommended actions. Questions were directed to (512) 873-0033.
Nationwide Distribution.
7,000