FDA Recall Terminated

VIA Medical Collection Bag; The collection bag is an accessory waste bag used in conjunction with the sterile LVM blood gas sensor with the LVM Monitor. The collection bag is packaged in a tyvek pouch which is included in either replacement kit (777-3103) or LVM sensor kit (777-3102).

Recall: Z-1531-2013 · Initiated May 24, 2013

Recall

Recall Number
Z-1531-2013
Event Number
65341
Firm
International Biomedical, Ltd.
FEI Number
1000058335
Product Code
LZF
Status
Terminated
Root Cause
Device Design
Initiated
May 24, 2013
Posted
June 11, 2013
Terminated
July 29, 2013
Address
8508 Cross Park Dr, Austin, TX, 78754-4532

Description

VIA Medical Collection Bag; The collection bag is an accessory waste bag used in conjunction with the sterile LVM blood gas sensor with the LVM Monitor. The collection bag is packaged in a tyvek pouch which is included in either replacement kit (777-3103) or LVM sensor kit (777-3102).

Reason

International Biomedical has received reports of interference between the collection bag and the extension set connectors. Evaluation of the affected components indicates that the collection bag connector can become occluded due to an over-tightening of the luer lock by the end user. To alleviate the occurrence of this issue in the fixture, the connector on the collection bag has been improved to

Action

International Biomedical decided to recall and sent "Product Bulletin" notification letters to their consignees on 05/24/2013 or 05/30/2013. The letter described the product issue and provided customers with recommended actions. Questions were directed to (512) 873-0033.

Distribution

Nationwide Distribution.

Quantity

7,000