FDA Recall Terminated

Baxter Evacuated Container, 500 mL, Sterile, non pyrogenic, Rx Only, Baxter Healthcare Corporation, Deerfield, IL 60015 USA, Made in USA, product code 1A8503.

Recall: Z-1530-2010 · Initiated February 15, 2010

Recall

Recall Number
Z-1530-2010
Event Number
55016
Firm
Baxter Healthcare Corp. Rt.
FEI Number
1416980
Product Code
KPE
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
February 15, 2010
Posted
April 29, 2010
Terminated
January 29, 2012
Address
120 & Wilson Rd, Round Lake, IL, 60073

Description

Baxter Evacuated Container, 500 mL, Sterile, non pyrogenic, Rx Only, Baxter Healthcare Corporation, Deerfield, IL 60015 USA, Made in USA, product code 1A8503.

Reason

Routine stability testing of the evacuated container product, along with trending of results, indicates that the pH will exceed the upper specification limit prior to the expiration date of this product.

Action

Urgent Product Recall letters dated 2/15/10 were mailed on the same date via first class mail to the affected direct accounts to the attention of the Director of Pharmacy, the Director of Materials Management, and the Director of Risk Management. The accounts were informed of the potential for the pH of the products to exceed the upper specification limits for pH prior to the product expiration date. The accounts were requested discontinue use and segregate the affected lots of products from inventory and call Baxter Healthcare Center for Service at 1-888-229-0001, Monday through Friday, 7:00 AM to 6:00 PM CT, to arrange for the return of the affected product for credit. The accounts were also requested to complete and fax back to Baxter the enclosed reply sheet confirming receipt of the letter and the amount of each product being returned. If the accounts further distributed the products, they were requested to forward the communication to those accounts. Questions regarding the communication are directed to the Medical Information Services at Baxter at 1-800-933-0303.

Distribution

United States -- Nationwide and Puerto Rico, Canada and Saudi Arabia

Quantity

79,308 units