FDA Recall Open, Classified

Programmers - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. 2090 Carelink 2. 29901 Encore

Recall: Z-1527-2021 · Initiated April 6, 2021

Recall

Recall Number
Z-1527-2021
Event Number
87662
Firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FEI Number
2182208
Product Code
NIK
Status
Open, Classified
Root Cause
Software design
Initiated
April 6, 2021
Address
8200 Coral Sea St Ne, Mounds View, MN, 55112-4391

Description

Programmers - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. 2090 Carelink 2. 29901 Encore

Reason

A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

Action

Beginning 06-APR-2021, Medtronic notified risk managers and practice managers of the field corrective action notification via 2- day UPS delivery of the consignee letter. Consignees were asked to confirm receipt of the notification. Medtronic Field Representatives have been instructed to begin installation of the software update. Outside the United States (OUS), communication began on 06- APR-2021 using locally approved methods. Confirmation will be obtained per locally approved methods. Medtronic Field Representatives will begin installation of the software update as it is available locally. A courtesy notification will be provided to physicians with affected devices.

Distribution

Worldwide distribution.

Quantity

37690 units