FDA Recall Terminated

Electrophysiology Imaging System

Recall: Z-1527-05 · Initiated April 28, 2005

Recall

Recall Number
Z-1527-05
Event Number
31930
Firm
Fischer Imaging Corporation
FEI Number
1418957
Product Code
JAA
Status
Terminated
Root Cause
Other
Initiated
April 28, 2005
Posted
September 9, 2005
Terminated
March 13, 2012
Address
12300 Grant St, Denver, CO, 80241-3120

Description

Electrophysiology Imaging System

Reason

Belt driving the Image Intensifier may break, causing the Intensifier to slip down, possibly causing injury to patient.

Action

Consignees were notified by certified letter on 04/27/2005.

Distribution

Nationwide, including VA facilities in GA and TX. Foreign distribution to Australia, Mexico and Poland. No military distribution.

Quantity