FDA Recall Terminated

Advance 35LP Low-Profile PTA Balloon Dilatation Catheter; Catalog numbers: PTA5-35-135-6-2 .0 and PTA5-35-135-4-4 .0; UDI: 10827002522647 and 10827002522524 Product Usage: The Advance 35LP Low Profile PTA Balloon Dilation Catheter is indicated for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Recall: Z-1521-2018 · Initiated April 16, 2018

Recall

Recall Number
Z-1521-2018
Event Number
79861
Firm
Cook Inc.
FEI Number
1820334
Product Code
DQY
Status
Terminated
Root Cause
Labeling mix-ups
Initiated
April 16, 2018
Terminated
June 27, 2019
Address
750 N Daniels Way, Bloomington, IN, 47404-9120

Description

Advance 35LP Low-Profile PTA Balloon Dilatation Catheter; Catalog numbers: PTA5-35-135-6-2 .0 and PTA5-35-135-4-4 .0; UDI: 10827002522647 and 10827002522524 Product Usage: The Advance 35LP Low Profile PTA Balloon Dilation Catheter is indicated for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Reason

Incorrect product labeling. Product labeled as 6mm x 2cm balloon are packaged with a 4mm x 4 cm balloon label and vice-versa.

Action

On April 16, 2018 Cook Medical Inc. mailed an Urgent Medical Device recall notification to affected customers. Customers were instructed to: 1) Examine your inventory immediately to determine if you have affected product(s), and quarantine affected product(s). Immediately cease all distribution and use of these products; 2) Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. NOTE: Unaffected products that are returned will not be credited; 3) Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return it via fax (812.339.7316) or email ([email protected]); 4) Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time) or by email to [email protected] Cook Incorporated

Distribution

Worldwide Distribution - US Nationwide in the states of AZ, DE, FL, GA, ID, IL, KY, MI, MO, NH, NJ, PA, RI, TX, UT and the country of Peru

Quantity

40 total products