Siemens SOMATOM Definition, SOMATOM Definition AS, SOMATOM Definition Flash and SOMATOM Definition Edge; Intended to produce cross-sectional images of the body. Computed tomography (CT) x-ray system.
Recall
- Recall Number
- Z-1521-2016
- Event Number
- 73593
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- March 2, 2016
- Posted
- April 19, 2016
- Terminated
- December 13, 2016
- Address
- 40 Liberty Blvd, Malvern, PA, 19355-1418
Description
Siemens SOMATOM Definition, SOMATOM Definition AS, SOMATOM Definition Flash and SOMATOM Definition Edge; Intended to produce cross-sectional images of the body. Computed tomography (CT) x-ray system.
Software bugs in these SOMATOM systems could possibly cause scan aborts, rescans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of additional contrast medium, accidental x-radiation to the user or impede customer workflow.
Siemens initiated a Customer Advisory Notice on 03/07/2016 to inform customers about actions for bug-fixing the following systems: SOMATOM Definition AS, SOMATOM Definition Flash and SOMATOM Definition Edge, which could possibly cause scan aborts, res cans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of additional contrast medium. Siemens mailed certified as well as via email distribution a Customer Advisory Notice to their customers to inform them about corrective actions for bug-fixing the following systems running with software version VA44A_SP3a, VA44A_SP3b, VA44A_SP3c, correction of potential safety issues. The issues described will be corrected with software VA44_SP4, handled as update CT051/15/P. Following the installation of this update, the mentioned malfunctions will be corrected.
Nationwide Distribution-including DC, and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
1096