FDA Recall Terminated

Software Version 2.8 & Below on VITROS 5,1 FS Chemistry Systems, Catalog Number 6801375, Global Trade Item Number 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number 6801890, Global Trade Item Number 10758750001644; IVD. Intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides and VITROS Chemistry Products MicroTip Reagents.

Recall: Z-1521-2015 · Initiated March 10, 2015

Recall

Recall Number
Z-1521-2015
Event Number
70685
Firm
Ortho-Clinical Diagnostics
FEI Number
1000136573
Product Code
JJE
Status
Terminated
Root Cause
Software in the Use Environment
Initiated
March 10, 2015
Posted
April 25, 2015
Terminated
June 1, 2018
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101

Description

Software Version 2.8 & Below on VITROS 5,1 FS Chemistry Systems, Catalog Number 6801375, Global Trade Item Number 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number 6801890, Global Trade Item Number 10758750001644; IVD. Intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides and VITROS Chemistry Products MicroTip Reagents.

Reason

Software Anomaly; It is possible for the device to process samples with a cartridge other than the intended cartridge, potentially leading to erroneous patient results. If this anomaly occurred previously, an indication would be a series of results that were similar across multiple samples (i.e., results could be believable or outside of the Reportable Range for the intended assay).

Action

On 3/10/2015, URGENT PRODUCT CORRECTION NOTIFICATION letters (Ref. CL2015-047, dated 3/10/2015) were sent to users notifying them of the issue and advising them of the required actions. Customers are to complete and return the Confirmation of Receipt no later than March 23, 2015. The firm is currently working on a resolution and will issue a follow up notification upon the software's availability. For questions, please contact Customer Technical Services at 1-800-421-3311.

Distribution

Worldwide Distribution -- USA, Puerto Rico, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain, and Venezuela.

Quantity

VITROS 5.1 System: Domestic - 909; Foreign - 1250; 5,1 Refurbished: Domestic - 66 units, Foreign - 152