Applied Medicals Epix and Direct Drive Laparoscopic Graspers, Model Numbers C4130 and C4140. Used for grasping and manipulating tissue during general or laparoscopic surgery.
Recall
- Recall Number
- Z-1521-2013
- Event Number
- 65217
- Firm
- Applied Medical Resources Corp 22872 Avenida Empresa # 3 Rancho Santa
- FEI Number
- 2027111
- Product Code
- NWV
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- May 1, 2013
- Posted
- June 7, 2013
- Terminated
- February 6, 2014
- Address
- Margarita, CA, 92688-2650
Description
Applied Medicals Epix and Direct Drive Laparoscopic Graspers, Model Numbers C4130 and C4140. Used for grasping and manipulating tissue during general or laparoscopic surgery.
Applied Medical is conducting a voluntary recall on specific lot numbers of its Epix and Direct Drive Laparoscopic Graspers. When the ratchet trigger is activated, if a great enough force is applied, there is a potential for the trigger to fracture and the jaws to remain in the closed position. The likelihood of this situation to occur and result in permanent patient injury is highly unlikely;
Applied Medical notified customers of the recall beginning on May 13, 2013. The "URGENT: MEDICAL DEVICE RECALL" letter identified the product, problem, and actions to be taken by the customers. For questions regarding this notice, contact the firm at (949) 713-8041.
Worldwide Distribution: USA Nationwide and countries of: Argentina, Austria, Australia, Bahrain, Belgium, Canada, Chile, Columbia, Denmark, Ecuador, Finland, France, Germany, Greece, India, Ireland, Israel, Italy, Jordan, Kuwait, Malaysia, Lebanon, Lybia, Mexico, Netherlands, New Zealand, Pakistan, Peru, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, Spain, South Africa, Switzerland, Taiwan-Province of China,Turkey, United Arabs Emirates United Kingdom, Uruguay, and Venezula-Bolivarian Republic.
127,305 units