FDA Recall Terminated

Omnia Soft Tissue Punch is a dental instrument designed to remove with precision a circular section of oral soft tissue (e.g., gum, palatal mucosa) during dental surgery, typically to allow access to a site for drilling and dental/orthodontic implant placement without the use of a scalpel. Model: 32.Z2000.00, 32.Z2002.00, 32.Z2004.00, 32.Z2006.00, 32.Z2007.00, 32.Z2008.00, 32.Z2009.00, 32.Z2020.00, 32.Z2022.00, 32.Z2024.00, 32.Z2026.00, 32.Z2027.00, 32.Z2028.00, 32.Z2030.00, 32.Z2032.00, 32.Z2034.00, 32.Z2036.00, 32.Z2038.00, 32.Z2040.00, 72.T1950.00, 72.T1951.00, 72.T1952.00, 72.T1953.00

Recall: Z-1518-2021 · Initiated March 30, 2021

Recall

Recall Number
Z-1518-2021
Event Number
87680
Firm
Omnia SRL Via Fratelli Cairoli 5 Fidenza Italy
FEI Number
3004361903
Product Code
EME
Status
Terminated
Root Cause
Vendor change control
Initiated
March 30, 2021
Posted
April 29, 2021
Terminated
November 8, 2023

Description

Omnia Soft Tissue Punch is a dental instrument designed to remove with precision a circular section of oral soft tissue (e.g., gum, palatal mucosa) during dental surgery, typically to allow access to a site for drilling and dental/orthodontic implant placement without the use of a scalpel. Model: 32.Z2000.00, 32.Z2002.00, 32.Z2004.00, 32.Z2006.00, 32.Z2007.00, 32.Z2008.00, 32.Z2009.00, 32.Z2020.00, 32.Z2022.00, 32.Z2024.00, 32.Z2026.00, 32.Z2027.00, 32.Z2028.00, 32.Z2030.00, 32.Z2032.00, 32.Z2034.00, 32.Z2036.00, 32.Z2038.00, 32.Z2040.00, 72.T1950.00, 72.T1951.00, 72.T1952.00, 72.T1953.00

Reason

Omnia SRL received a notification from the contract manufacturer informing them of inadequate sterilization processes which have been taking place at their contracted sterilizing facility since 2016.

Action

On 03/30/2021, Omnia issued an Urgent Medical Device Removal notice via letter to customers notifying them Omnia has been made aware of a situation at a contract sterilization facility regarding inadequate or inappropriate sterilization processes being followed by employees of the contract sterilizer.

Distribution

Domestic: AK, IL, PA, OK, TX, UT, NY, CA. Foreign: Australia, Austria, Belgium, Bulgaria, Denmark, Dominican Republic, Estonia, France, Germany, Greece, Hungary, Ireland, Italy, Lebanon, Netherlands, New Zealand, Poland, Russia, Slovenia, Spain, Sweden, Switzerland, United Kingdom, Uzbekistan, Canada, Malta,

Quantity

276,901 units