FDA Recall Terminated

Raindrop Near Vision Inlay, Model# RD1-1

Recall: Z-1518-2017 · Initiated February 7, 2017

Recall

Recall Number
Z-1518-2017
Event Number
76422
Firm
Revision Optics Inc
FEI Number
3005956347
Product Code
LQE
Status
Terminated
Root Cause
Use error
Initiated
February 7, 2017
Posted
March 21, 2017
Terminated
June 21, 2017
Address
25651 Atlantic Ocean Dr, Ste A1, Lake Forest, CA, 92630-8835

Description

Raindrop Near Vision Inlay, Model# RD1-1

Reason

ReVision Optics has updated the instructions for use (IFU) for the Raindrop Near Vision Inlay to emphasize that only Balanced Salt Solution (BSS) may be used to irrigate under the flap and that topical medications or lubricants should not be administered until centration of the inlay and proper flap positioning has been confirmed at the slit lamp.

Action

A Dear Doctor letter was sent to inform customers that ReVision Optics wanted to share a surgical update regarding the Raindrop Near Vision Inlay procedure. Customers are informed that no ophthalmic medications or lubricants should be used immediately prior to or during placement of the Raindrop Near Vision Inlay. Only Balanced Salt Solution (BSS) should be used to irrigate under the flap during the Raindrop procedure. Customers are informed that ophthalmic medications or lubricants should be administered only after verifying proper inlay centration and flap positioning as observed at the slit lamp. This ensures that no further inlay or flap manipulation is required. ReVision Optics has updated the instructions for use (IFU) for the Raindrop Near Vision Inlay to emphasize that only BSS may be used to irrigate under the flap and that topical medications or lubricants should not be administered until centration of the inlay and proper flap positioning has been confirmed at the slit lamp. Customers with questions are instructed to contact Luis Vargas regarding the surgical procedure update.

Distribution

US: AL, AZ, CA, FL, GA, HI, IL, IN, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, PA, SD, TN, TX, UT, WA

Quantity

1,279 devices