FDA Recall Terminated

GE Healthcare Innova 4100 Digital Fluoroscopic Imaging System with Bolus Chasing option.

Recall: Z-1517-2008 · Initiated October 21, 2007

Recall

Recall Number
Z-1517-2008
Event Number
45538
Firm
GE Healthcare
FEI Number
2126677
Product Code
MQB
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
October 21, 2007
Posted
September 29, 2008
Terminated
December 17, 2011
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Healthcare Innova 4100 Digital Fluoroscopic Imaging System with Bolus Chasing option.

Reason

Incorrect dose data: GE Healthcare has found that after 6 days and 4 hours without performing a system reset or a system reboot (turning it off/on), the displayed dose data on these systems (Innova 3100 / Innova 4100 with Bolus option) may be underestimated by up to 50%. In order to avoid this issue, it is necessary to perform a system reset or reboot every day. GE Healthcare is working on a c

Action

GE sent all customers with affected units a letter on 11/15/07 titled Urgent Medical Device Correction. In the US this was sent by a traceable means (either US Postal Service with receipt acknowledgement (registered mail) or equivalent courier service such as FEDEX overnight Delivery). The letter was sent to 3 titles within the affected accounts; Hospital Administrator, Manager of Radiology/Cardiology and Radiologist/ Cardiologist. A correction letter identified a short term recommendation to perform a system reset or reboot every day and stated that a software correction will be installed at their site at no cost.

Distribution

Worldwide Distribution: USA, Argentina, Australia, Canada, China, Germany, Japan, Mexico and Russia.

Quantity

146 total