GE Healthcare Innova 4100 Digital Fluoroscopic Imaging System with Bolus Chasing option.
Recall
- Recall Number
- Z-1517-2008
- Event Number
- 45538
- Firm
- GE Healthcare
- FEI Number
- 2126677
- Product Code
- MQB
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- October 21, 2007
- Posted
- September 29, 2008
- Terminated
- December 17, 2011
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
GE Healthcare Innova 4100 Digital Fluoroscopic Imaging System with Bolus Chasing option.
Incorrect dose data: GE Healthcare has found that after 6 days and 4 hours without performing a system reset or a system reboot (turning it off/on), the displayed dose data on these systems (Innova 3100 / Innova 4100 with Bolus option) may be underestimated by up to 50%. In order to avoid this issue, it is necessary to perform a system reset or reboot every day. GE Healthcare is working on a c
GE sent all customers with affected units a letter on 11/15/07 titled Urgent Medical Device Correction. In the US this was sent by a traceable means (either US Postal Service with receipt acknowledgement (registered mail) or equivalent courier service such as FEDEX overnight Delivery). The letter was sent to 3 titles within the affected accounts; Hospital Administrator, Manager of Radiology/Cardiology and Radiologist/ Cardiologist. A correction letter identified a short term recommendation to perform a system reset or reboot every day and stated that a software correction will be installed at their site at no cost.
Worldwide Distribution: USA, Argentina, Australia, Canada, China, Germany, Japan, Mexico and Russia.
146 total