American Optisurgical's TX1 Tissue Removal System Console. The console is a software unit that is indicated for use in surgical procedures.
Recall
- Recall Number
- Z-1516-2013
- Event Number
- 65113
- Firm
- American Optisurgical Inc
- FEI Number
- 1000135560
- Product Code
- LFL
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- March 30, 2013
- Posted
- June 6, 2013
- Terminated
- March 24, 2014
- Address
- 26902 Vista Ter, Lake Forest, CA, 92630-8123
Description
American Optisurgical's TX1 Tissue Removal System Console. The console is a software unit that is indicated for use in surgical procedures.
The recall has been initiated due to an overheating protection voltage suppressor on the redundant safety protection circuit board. Secondary to the recall, American Optisurgical, Inc. will be making a correction to add spacers between the valve and cassette to correct a small irrigation leak that can occur after the priming cycle. Use of this product may cause the device to lose power and stop wo
"Urgent Medical Device Recall" notifications letters were sent to customers on May 20, 2013 via USPS registered mail. Included in the Recall Notification Letter were the recommended actions and instructions for customers. Questions were referred to (949) 580-1266.
Worldwide distribution, US, including the states of United States: Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Iowa, Illinois, Indiana, Kansas, Kentucky, Massachusetts, Maryland, Michigan, Minnesota, Missouri, North Carolina, Nebraska, New Jersey, New York, Ohio, Pennsylvania, South Carolina, Tennessee, Texas, Utah, Virginia, Washington Wisconsin, West Virginia and the countries of Singapore and the Philippines.
194 units