FDA Recall Terminated

American Optisurgical's TX1 Tissue Removal System Console. The console is a software unit that is indicated for use in surgical procedures.

Recall: Z-1516-2013 · Initiated March 30, 2013

Recall

Recall Number
Z-1516-2013
Event Number
65113
Firm
American Optisurgical Inc
FEI Number
1000135560
Product Code
LFL
Status
Terminated
Root Cause
Device Design
Initiated
March 30, 2013
Posted
June 6, 2013
Terminated
March 24, 2014
Address
26902 Vista Ter, Lake Forest, CA, 92630-8123

Description

American Optisurgical's TX1 Tissue Removal System Console. The console is a software unit that is indicated for use in surgical procedures.

Reason

The recall has been initiated due to an overheating protection voltage suppressor on the redundant safety protection circuit board. Secondary to the recall, American Optisurgical, Inc. will be making a correction to add spacers between the valve and cassette to correct a small irrigation leak that can occur after the priming cycle. Use of this product may cause the device to lose power and stop wo

Action

"Urgent Medical Device Recall" notifications letters were sent to customers on May 20, 2013 via USPS registered mail. Included in the Recall Notification Letter were the recommended actions and instructions for customers. Questions were referred to (949) 580-1266.

Distribution

Worldwide distribution, US, including the states of United States: Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Iowa, Illinois, Indiana, Kansas, Kentucky, Massachusetts, Maryland, Michigan, Minnesota, Missouri, North Carolina, Nebraska, New Jersey, New York, Ohio, Pennsylvania, South Carolina, Tennessee, Texas, Utah, Virginia, Washington Wisconsin, West Virginia and the countries of Singapore and the Philippines.

Quantity

194 units