FDA Recall Terminated

ADVIA Centaur CP System, Automated Immunoassay Analyzer, Part /Catalogue Number 086-A001.

Recall: Z-1516-06 · Initiated August 4, 2006

Recall

Recall Number
Z-1516-06
Event Number
36181
Firm
Siemens Healthcare Diagnostics
FEI Number
2432235
Product Code
JJE
Status
Terminated
Root Cause
Software design
Initiated
August 4, 2006
Posted
September 23, 2006
Terminated
July 30, 2012
Address
511 Benedict Ave, Tarrytown, NY, 10591-5005

Description

ADVIA Centaur CP System, Automated Immunoassay Analyzer, Part /Catalogue Number 086-A001.

Reason

Sample/Patient mis-identification (software defect)-the system can associate a test result and sample identification (SID) with an incorrect patient name when the Patient Demographics feature is used and the Patient Identification (PID) field is left blank

Action

On 8/04/06, a Support Bulletin was e-mailed to all affected Bayer HealthCare LLC Branches worldwide in order to inform them of the issue and to provide them with instructions for this 'field correction' until the new software version is issued. A Customer Bulletin and Confirmation Form (FaxBack) were prepared and sent to the affected Bayer Healthcare LLC Branches on 8/04/06 for communication with the affected customers worldwide.

Distribution

Worldwide

Quantity

Domestic - 101, Foreign - 153