ADVIA Centaur CP System, Automated Immunoassay Analyzer, Part /Catalogue Number 086-A001.
Recall
- Recall Number
- Z-1516-06
- Event Number
- 36181
- Firm
- Siemens Healthcare Diagnostics
- FEI Number
- 2432235
- Product Code
- JJE
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- August 4, 2006
- Posted
- September 23, 2006
- Terminated
- July 30, 2012
- Address
- 511 Benedict Ave, Tarrytown, NY, 10591-5005
Description
ADVIA Centaur CP System, Automated Immunoassay Analyzer, Part /Catalogue Number 086-A001.
Sample/Patient mis-identification (software defect)-the system can associate a test result and sample identification (SID) with an incorrect patient name when the Patient Demographics feature is used and the Patient Identification (PID) field is left blank
On 8/04/06, a Support Bulletin was e-mailed to all affected Bayer HealthCare LLC Branches worldwide in order to inform them of the issue and to provide them with instructions for this 'field correction' until the new software version is issued. A Customer Bulletin and Confirmation Form (FaxBack) were prepared and sent to the affected Bayer Healthcare LLC Branches on 8/04/06 for communication with the affected customers worldwide.
Worldwide
Domestic - 101, Foreign - 153