FDA Recall Terminated

Sony Medical Monitor LMD-1951MD 19 inch

Recall: Z-1514-2015 · Initiated March 5, 2015

Recall

Recall Number
Z-1514-2015
Event Number
70702
Firm
Sony Latin America
FEI Number
3010052497
Product Code
KQM
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
March 5, 2015
Posted
April 23, 2015
Terminated
December 14, 2016
Address
5201 Blue Lagoon Dr, Miami, FL, 33126-2064

Description

Sony Medical Monitor LMD-1951MD 19 inch

Reason

Monitor has either not turned on or has lost image during clinical use.

Action

SOLA sent an URGENT MEDICAL DEVICE FIELD ACTION dated March 2, 2015, to all affected customers. Within that notification SOLA requests that you send the customer letter included in Attachment 1 of this notification to all customers who have purchased the affected monitors from you. The letter instructs customers who already possess an AC adapter for each of their affected monitors to only use the AC adaptor when the monitors are in use. If you have units in inventory that can be provided with an AC adapter, we will provide under separate cover kits that contain: 1) a sticker warning and 2) an adhesive seal to go over the AC outlet and 3) correction instructions document. If you have units in inventory that cannot be accompanied by an AC adapter, for each of these affected monitors, do not ship the unit until the permanent corrective action described below has been applied. Please complete the MEDICAL DEVICE RECALL RETURN RESPONSE included in Attachment 2 to the contact person identified on the form. Please contact me below if you have any questions. 305-260-4000 Hours of operation: 8am to 8pm (Monday through Friday, excluding holidays).

Distribution

Worldwide Distribution - US Distribution to the states of : FL and NC., and to the countries of : Ecuador, Costa Rica, Brazil, Nicaragua, Venezuela, Peru, Colombia, Dominican Republic, and South Korea.

Quantity

4,764 units