FDA Recall Terminated

HexaPOD evo RT Couchtop. To support and aid in positioning a patient during radiation therapy.

Recall: Z-1514-2014 · Initiated April 17, 2014

Recall

Recall Number
Z-1514-2014
Event Number
68088
Firm
Elekta, Inc.
FEI Number
1037831
Product Code
JAI
Status
Terminated
Root Cause
Component design/selection
Initiated
April 17, 2014
Posted
April 25, 2014
Terminated
January 29, 2015
Address
400 Perimeter Center Ter Ne, Ste 50, Atlanta, GA, 30346-1227

Description

HexaPOD evo RT Couchtop. To support and aid in positioning a patient during radiation therapy.

Reason

If screws are not properly tightened with a torque wrench they may become loose over time.

Action

Corrective Action #1 - Important Field Safety Notice 618 01 302 002 was issued April 2014. The Notice informs end users about the problem and the potential clinical impact. It also give a recommendation how the risk can be mitigated - inspect the screws and if loose contact Elekta Service to tighten the screws with 20 Nm. The notice also advises the customer that a modification kit will be released that introduces an improved screw solution. The solution will be implemented on all affected products when available. Corrective Action #2 Permanent Solution - Replace Current Screws - An improved screw has been designed and will be available for Elekta Field Service organization to replace on all affected products at the end of April. The new design solution provides a very long used thread depth, exceeding the standard recommended thread-depths by far. This design will prevent the screws from loosening.

Distribution

Worldwide Distribution-USA including DC and the states of AZ, CA, CT, FL, GA, IL, HI, KY, MD, MA, MI, MS, MO, NJ, NY, NH, NC, ND, OH, OK, OR, PA, SC, TX, VA, WA, WI, and WY, and the countries of Algeria, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kazakhstan, Mexico, Netherlands, New Zealand, Peru, Poland, Portugal, Russia, Saudi Arabia, South Korea, Span, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.

Quantity

266