FDA Recall Terminated

CSA Medical truFreeze Console- Cryosurgical Unit Cryogenic Surgical Device Model: CC3-01

Recall: Z-1513-2015 · Initiated March 25, 2015

Recall

Recall Number
Z-1513-2015
Event Number
70851
FEI Number
3010140265
Product Code
GEH
Status
Terminated
Root Cause
Software design
Initiated
March 25, 2015
Posted
April 23, 2015
Terminated
August 21, 2015
Address
CSA Medical 91 Hartwell Ave, Lexington, MA, 02421-3137

Description

CSA Medical truFreeze Console- Cryosurgical Unit Cryogenic Surgical Device Model: CC3-01

Reason

TruFreeze Console caused a higher rate of liquid nitrogen (cryogen) to be delivered and may cause: stricture,scarring, bradycardia, or pneumothorax

Action

CSA Medical issued letter dated 3/25/15 advising users of the problem. Users provided with: The mitigations available to the active venting procedures coupled with the extremely unlikely probability of injury, rare risk, thus allowing the physician to continue with active venting procedures. In regards a passive venting procedure may not have sufficient mitigation to allow the physician to identify the potential hazard with sufficient time to preclude a potential for injury. Therefore, passive users will be instructed not to use the system until a software improvement is put into place. Additionally, no catheters or consoles will be shipped to passive venting users of truFreeze. A response form to be signed and returned confirming receipt of the notification.Questions contact: Stephen Mascioli, MD 781-538-4755 [email protected].

Distribution

Nationwide

Quantity

82 units