CSA Medical truFreeze Console- Cryosurgical Unit Cryogenic Surgical Device Model: CC3-01
Recall
- Recall Number
- Z-1513-2015
- Event Number
- 70851
- FEI Number
- 3010140265
- Product Code
- GEH
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- March 25, 2015
- Posted
- April 23, 2015
- Terminated
- August 21, 2015
- Address
- CSA Medical 91 Hartwell Ave, Lexington, MA, 02421-3137
Description
CSA Medical truFreeze Console- Cryosurgical Unit Cryogenic Surgical Device Model: CC3-01
TruFreeze Console caused a higher rate of liquid nitrogen (cryogen) to be delivered and may cause: stricture,scarring, bradycardia, or pneumothorax
CSA Medical issued letter dated 3/25/15 advising users of the problem. Users provided with: The mitigations available to the active venting procedures coupled with the extremely unlikely probability of injury, rare risk, thus allowing the physician to continue with active venting procedures. In regards a passive venting procedure may not have sufficient mitigation to allow the physician to identify the potential hazard with sufficient time to preclude a potential for injury. Therefore, passive users will be instructed not to use the system until a software improvement is put into place. Additionally, no catheters or consoles will be shipped to passive venting users of truFreeze. A response form to be signed and returned confirming receipt of the notification.Questions contact: Stephen Mascioli, MD 781-538-4755 [email protected].
Nationwide
82 units