Drger Perseus A500 Anesthesia Workstation

Recall: Z-1512-2024 · Initiated March 20, 2024

Recall

Recall Number
Z-1512-2024
Event Number
94269
Firm
Draeger Medical, Inc.
FEI Number
2517967
Product Code
BSZ
Status
Open, Classified
Root Cause
Labeling design
Initiated
March 20, 2024
Posted
April 18, 2024
Address
3135 Quarry Rd, Telford, PA, 18969-1042

Description

Drger Perseus A500 Anesthesia Workstation

Reason

Unexpected shutdown while operating on battery power.

Action

Consignees were sent an Urgent Medical Device Recall notification by mail on 3/20/24. The notice informs consignees of the possibility of device shutdown with possible loss of ventilation, which consignees are instructed to either use an emergency ventilation bag or apply manual ventilation with the Perseus with the provided instructions. Consignees are to perform a battery test on all affected devices in their inventory per the provided instructions in the recall notification. If the tested unit(s) fails the battery test, consignees are to contact Drager Service Technical Support between 8:00 AM and 8:00 PM at 1-800-437-2437 (option 2, option 2, option 2) to arrange replacement of the device's battery. Consignees are also instructed to avoid short discharging/charging cycles of devices; if this cannot be accommodated device batteries are to be tested every three months. A supplemental IFU has been released to include the additional maintenance instructions outlined in the recall notification and can be requested from the firm. Consignees with any questions can contact Michael Kelhart between 8:00 AM to 4:30 PM EST at 267-664-1131 or [email protected].

Distribution

Domestic: Nationwide Distribution; Foreign: Angola, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-se- S, Bulgaria, Canada, Chili, China, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polynesia, French Guiana, Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Liechtenstein, Lithuania, Malaysia, Maldives, Mauritius, Mayotte, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, Uganda, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.

Quantity

16,841 units