Boston Scientific, LATITUDE Programming System, Model 3300 - Product Usage:The Boston Scientific LATITUDE Programming System is intended for use in hospital and clinical environments to communicate with Boston Scientific implantable systems. The software in use controls all communication functions for the Pulse Generator (PG). Detailed software applications instructions can be found in the associated product literature for the PG being interrogated.
Recall
- Recall Number
- Z-1511-2020
- Event Number
- 85077
- Firm
- Boston Scientific Corporation
- FEI Number
- 2124215
- Product Code
- LWS
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- February 12, 2020
- Posted
- March 3, 2020
- Terminated
- April 14, 2023
- Address
- 4100 Hamline Ave N, Bldg 3, Saint Paul, MN, 55112-5700
Description
Boston Scientific, LATITUDE Programming System, Model 3300 - Product Usage:The Boston Scientific LATITUDE Programming System is intended for use in hospital and clinical environments to communicate with Boston Scientific implantable systems. The software in use controls all communication functions for the Pulse Generator (PG). Detailed software applications instructions can be found in the associated product literature for the PG being interrogated.
Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update Application then directed these programmers to download software applications that have not been approved by the US Food and Drug Administration.
The firm initiated the recall by letter on 12/12/2020. The letter explained the error and requested the consignee segregate the units for removal. The firm will provide another model of programmer to be used in the interim.
US Nationwide distribution in the states of NC, TN, and MO.
4 devices