TS ROC Crown Component 4UR (5/pk), TS ROC Crown Component 5UR (5/pk), TS ROC Crown Component 6UR (5/pk), TS ROC Crown Component 7UR (5/pk), TS ROC Crown Component 8UR (5/pk), TS ROC Crown Component 4LR (5/pk), TS ROC Crown Component 5LR (5/pk), TS ROC Crown Component 6LR (5/pk), TS ROC Crown Component 7LR (5/pk), TS ROC Crown Component 8LR (5/pk), TS ROC Crown Component 4UL (5/pk), TS ROC Crown Component 5UL (5/pk), TS ROC Crown Component 6UL (5/pk), TS ROC Crown Component 7UL (5/pk), TS ROC Crown Component 8UL (5/pk), TS ROC Crown Component 4LL (5/pk), TS ROC Crown Component 5LL (5/pk), TS ROC Crown Component 6LL (5/pk), TS ROC Crown Component 7LL (5/pk), TS ROC Crown Component 8LL (5/pk) A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance.
Recall
- Recall Number
- Z-1511-2015
- Event Number
- 70787
- Firm
- Specialty Appliance Works, Inc.
- FEI Number
- 3004158247
- Product Code
- ECN
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- March 9, 2015
- Posted
- April 22, 2015
- Terminated
- January 12, 2017
- Address
- 4905 Hammond Industrial Dr, Cumming, GA, 30041-3914
Description
TS ROC Crown Component 4UR (5/pk), TS ROC Crown Component 5UR (5/pk), TS ROC Crown Component 6UR (5/pk), TS ROC Crown Component 7UR (5/pk), TS ROC Crown Component 8UR (5/pk), TS ROC Crown Component 4LR (5/pk), TS ROC Crown Component 5LR (5/pk), TS ROC Crown Component 6LR (5/pk), TS ROC Crown Component 7LR (5/pk), TS ROC Crown Component 8LR (5/pk), TS ROC Crown Component 4UL (5/pk), TS ROC Crown Component 5UL (5/pk), TS ROC Crown Component 6UL (5/pk), TS ROC Crown Component 7UL (5/pk), TS ROC Crown Component 8UL (5/pk), TS ROC Crown Component 4LL (5/pk), TS ROC Crown Component 5LL (5/pk), TS ROC Crown Component 6LL (5/pk), TS ROC Crown Component 7LL (5/pk), TS ROC Crown Component 8LL (5/pk) A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance.
The Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS.
Specialty Applicances sent an Urgent Voluntary Productionn Field Correction Notification dated March 9, 2015, to all affected customers who received the Herbst Appliances and for sales supply orders. The letter advised doctors to monitor the appliance during routine follow-up appointments for signs of corrosion, and for sales supply orders to return affected components to the lab for replacement. Customers were asked to complete and return the enclosed Field Correctivve Action Response Form and return via fax to 678-513-7345. To return affected devices call 800-522-4636 to receive instruction as to how to return it to Speciality Applicances. In the event that customers observe signs of corrosion or a fracture, they should contact customer service at 800-522-4636 or their sales representative. For questions customers should call 800-522-4636 For questions regarding this recall call 678-513-4408, ext 226.
Worldwide Distribution - US nationwide including Puerto Rico, Brazil, China, Colombia, Germany, Ireland, Mexico, Switzerland, and United Kingdom.
14,428 Universal Nuts