FDA Recall Open, Classified

Boppli: Bedside Device Kit, REF: 54-0002; Sensor Band, REF: 51-0007

Recall: Z-1508-2025 · Initiated February 24, 2025

Recall

Recall Number
Z-1508-2025
Event Number
96414
Firm
PYRAMES INC
FEI Number
3014466900
Product Code
DXN
Status
Open, Classified
Root Cause
Software design
Initiated
February 24, 2025
Posted
April 2, 2025
Address
21730 Stevns Crk Blvd, Ste 201a, Cupertino, CA, 95014-1171

Description

Boppli: Bedside Device Kit, REF: 54-0002; Sensor Band, REF: 51-0007

Reason

Single-use battery-powered blood pressure monitor, for use on neonates and infants, due to vibratory noise, may not adequately detect changes in blood pressure per specification. Containment actions: 1) Ventilator lines not to touch the isolette (e.g. mattress, side rails, pillow) or patient, 2) Move affected monitor to patient's foot.

Action

The firm sent a MEDICAL DEVICE CORRECTIONS NOTTICE" dated 3/5/2025 to its customers describing the product, problem and actions to be taken. Starting on 2/24/2025, the firm started providing training to users to inform them of the following: Signal may be green, but high frequency noise appears superimposed on pulse waveform signal. Containment actions: 1) Ensure that ventilator lines do not physically touch the isolette (e.g. mattress, side rails, pillow) or baby. 2) Move Boppli to patient's foot. On 3/7/2025, correction notices were emailed to customers who in addition to the above were asked to do the following: 3) Use cuff measurement(s) to confirm that the signal is no longer affected by the high-frequency oscillatory ventilation (HFOV) interference. 4) Share this notice with anyone who needs to be made aware within the organization and forward to any organization where potentially affected devices have been transferred. 5) If these actions do not mitigate the issue, discontinue using the affected blood pressure monitoring device. Firm will be releasing a new version of software that will recognize the HFOV interference, indicate its presence by changing the signal color to red, and exclude that data from BP calculations. It is expected that this solution will be implemented by April 30, 2025. The Instructions for Use (55-0001 Rev. E) was updated to include warnings of vibratory interference, what it looks like and what to do when it is identified. Please contact [email protected] or 408-569-5215 for further information.

Distribution

US distribution to states of: MA and CA

Quantity

35