FDA Recall Open, Classified

Cardiosave Hybrid IABP - Product Usage: used to inflate and deflate intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation, Model Numbers 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65

Recall: Z-1506-2021 · Initiated March 22, 2021

Recall

Recall Number
Z-1506-2021
Event Number
87482
Firm
Datascope Corp.
FEI Number
3001418283
Product Code
DSP
Status
Open, Classified
Root Cause
Software design
Initiated
March 22, 2021
Posted
April 28, 2021
Address
1300 Macarthur Blvd, Mahwah, NJ, 07430-2052

Description

Cardiosave Hybrid IABP - Product Usage: used to inflate and deflate intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation, Model Numbers 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65

Reason

There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS).

Action

URGENT MEDICAL DEVICE CORRECTION notification letters dated 3/22/21 were distributed to customers. PLEASE FORWARD THIS INFORMATION TO ALL CURRENT AND POTENTIAL CARDIOSAVE HYBRID and CARDIOSAVE RESCUE IABP USERS WITHIN YOUR HOSPITAL / FACILITY. IF YOU ARE A DISTRIBUTOR WHO HAS SHIPPED ANY AFFECTED PRODUCTS TO CUSTOMERS, PLEASE FORWARD THIS DOCUMENT TO THEIR ATTENTION FOR APPROPRIATE ACTION. Corrective Action: Datascope/Getinge is currently developing a software correction to address this issue. A Datascope/Getinge service representative will contact you to schedule the installation of the updated software. This work will be done at no cost to your facility. Please complete and sign the attached MEDICAL DEVICE CORRECTION RESPONSE FORM (page 5) to acknowledge that you have received this notification. Return the completed form to Datascope/Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to 1-800-506-1644.

Distribution

Worldwide distribution - US Nationwide distribution.

Quantity

10087 total