FDA Recall Open, Classified

Scooter

Recall: Z-1505-2021 · Initiated April 1, 2021

Recall

Recall Number
Z-1505-2021
Event Number
87673
Firm
Merits Holdings Co. dba Merits Health Products, Inc.
FEI Number
1000221992
Product Code
INI
Status
Open, Classified
Root Cause
Process change control
Initiated
April 1, 2021
Address
4245 Evans Ave, Fort Myers, FL, 33901-9311

Description

Scooter

Reason

Seat baseplate pin has broken in the field.

Action

A Recall notification letter titled, "MEDICAL DEVICE FIELD CORRECTION" was sent to consignees on 04/07/2021 via email and by certified mail on 04/16/2021. The firm stated they made phone call notifications on 04/01/2021. The letter instructs the consignee to Contact impacted customer using a personalized customer letter (template provided). The notification can be sent by mail, email or phone. (IF by phone, please keep a log of call. Date/Time/Result); replace the old seat plate with new one provided using the SOP provided. The field correction should take approximately 15 minutes to complete.; complete and return the Service and Repair Report (sent with new part), inspect the installation and note it on the appropriate Scooter or Power Chair Inspection Record provided; use the RA shipping label (sent with new part) and return old seat plate to Merits Health Products along with the Service and Repair Report (one filled out for each unit); and when all of the units in the possession of the consignee are finished, please return the completed Inspection Report/s to Merits by mail or email.

Distribution

Distributed nationwide to AL, AK, AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, IO, KS, KY, LA, MN, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, PR, SC, TN, TX, UT, VT, VA, WA, WV, WI and internationally to Canada

Quantity

189 units