FDA Recall Open, Classified

Atellica IM Cortisol 50T (Material Number 10995538), 250T (Material Number 10995537), and REF 250T (Material Number 11206248). In vitro diagnostic use in the quantitative determination of cortisol in human serum, plasma (EDTA and lithium heparin), and urine.

Recall: Z-1504-2023 · Initiated March 16, 2023

Recall

Recall Number
Z-1504-2023
Event Number
92059
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2432235
Product Code
JFT
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
March 16, 2023
Posted
May 3, 2023
Address
511 Benedict Ave, Tarrytown, NY, 10591-5005

Description

Atellica IM Cortisol 50T (Material Number 10995538), 250T (Material Number 10995537), and REF 250T (Material Number 11206248). In vitro diagnostic use in the quantitative determination of cortisol in human serum, plasma (EDTA and lithium heparin), and urine.

Reason

There is a negative bias with urine patient samples and urine cortisol Quality Control (QC) results intermittently out of range low.

Action

An Urgent Medical Device Correction notification letter dated 3/17/23 was sent to customers. Actions to be Taken by the Customer Discontinue use of urine patient samples with the Atellica IM Cortisol and ADVIA Centaur Cortisol assays until further notice. Customers may continue to use serum and plasma sample types with the Atellica IM Cortisol and ADVIA Centaur Cortisol assays. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. An additional Resolution Customer Notification letter dated 2/29/24 was sent to customers explaining an improvement to raw material specification that restores assay performance with urine samples. Customers are asked to follow the instructions for urine patient samples and urine cortisol QC as stated in the notification and if using products listed in the original field action notice continue following the guidance as stated in the Field Action Letter. Customers with any questions are to contact the Customer Care Center or local Siemens Healthineers technical support representative.

Distribution

Worldwide distribution - US Nationwide.

Quantity

34,746 kits