FDA Recall Terminated

Caridian BCT COBE Spectra System Blood Warmer, Catalog number 950000000.

Recall: Z-1504-2012 · Initiated October 26, 2011

Recall

Recall Number
Z-1504-2012
Event Number
53828
Firm
CaridianBCT, Inc.
FEI Number
1000117351
Product Code
LKN
Status
Terminated
Root Cause
Device Design
Initiated
October 26, 2011
Posted
May 9, 2012
Terminated
May 11, 2012
Address
10811 W Collins Ave, Lakewood, CO, 80215-4440

Description

Caridian BCT COBE Spectra System Blood Warmer, Catalog number 950000000.

Reason

Potential for air entry into the fluid pathway of the COBE Spectra due to a leak at the return luer connection to an elevated bloodwarmer when directions for use are not followed.

Action

Caridian BCT sent a letter to customers dated June 2011. It was a Field labeling addendum to the COBE Spectra Apheresis System Operator's Manual stating: "When connecting a blood warmer tubing set to the return line, ensure that the tubing connection is tight. Put the luer connection no higher than 20 in (50 cm) above the return access to prevent the possibility of air entering the tubing." This correction was distributed to all consignees and is included with the current version of the owner's manual.

Distribution

Worldwide distribution including USA, Canada, Algeria, Argentina, Australia, Austria, Bahamas, Belgium, Brazil, Chile, China, Colombia, Czech Republic, Denmark, Netherlands, UK, Hong Kong, Hungary, India, Italy, Iran, Ireland, Israel, Jordan, Korea, Kuwait, Lebanon, Libya, Luxembourg, Egypt, Finland, France, Germany, Japan, Malaysia, Mexico, Norway, Oman, Pakistan, Philippines, Poland, Portugal, Romania, Russia, Spain, Saudi Arabia, South Africa, Sweden, Switzerland, Taiwan, Thailand, Ukraine, Uruguay, Vatican, Vietnam, and Greece.

Quantity

2459 machines