FDA Recall Terminated

LIFEPAK 20 external defibrillator/monitor

Recall: Z-1502-06 · Initiated August 30, 2006

Recall

Recall Number
Z-1502-06
Event Number
36190
Firm
Medtronic Emergency Response Systems, Inc.
FEI Number
3015876
Product Code
LDD
Status
Terminated
Root Cause
Other
Initiated
August 30, 2006
Posted
September 13, 2006
Terminated
March 25, 2011
Address
11811 Willows Rd Ne, Redmond, WA, 98052-2003

Description

LIFEPAK 20 external defibrillator/monitor

Reason

Devices with v38 system software do not display a ''LOW BATTERY: CONNECT TO AC POWER'' message when the monitor is on backup (DC) battery power and may shut down without warning.

Action

On August 30, 2006 the firm issued a letter dated August 2006 and titled MEDICAL DEVICE CORRECTION. The letter advises of the issue, explains how to resume power if power is shut down without warning, and advsies that a service representative will contact the consignee to make arrangements to install new software. On 10/19/06 the firm mailed an identical letter dated October 2006 to customers for units that were later identified as being subject to the recall.

Distribution

Units were distributed worldwide to hospitals and medical centers.

Quantity

2,831 worldwide