LIFEPAK 20 external defibrillator/monitor
Recall
- Recall Number
- Z-1502-06
- Event Number
- 36190
- Firm
- Medtronic Emergency Response Systems, Inc.
- FEI Number
- 3015876
- Product Code
- LDD
- Status
- Terminated
- Root Cause
- Other
- Initiated
- August 30, 2006
- Posted
- September 13, 2006
- Terminated
- March 25, 2011
- Address
- 11811 Willows Rd Ne, Redmond, WA, 98052-2003
Description
LIFEPAK 20 external defibrillator/monitor
Devices with v38 system software do not display a ''LOW BATTERY: CONNECT TO AC POWER'' message when the monitor is on backup (DC) battery power and may shut down without warning.
On August 30, 2006 the firm issued a letter dated August 2006 and titled MEDICAL DEVICE CORRECTION. The letter advises of the issue, explains how to resume power if power is shut down without warning, and advsies that a service representative will contact the consignee to make arrangements to install new software. On 10/19/06 the firm mailed an identical letter dated October 2006 to customers for units that were later identified as being subject to the recall.
Units were distributed worldwide to hospitals and medical centers.
2,831 worldwide