FDA Recall Terminated

Turbett Surgical Container, TS1500

Recall: Z-1501-2022 · Initiated April 14, 2022

Recall

Recall Number
Z-1501-2022
Event Number
90578
Firm
Turbett Surgical, Inc.
FEI Number
3012421755
Product Code
FRG
Status
Terminated
Root Cause
Process control
Initiated
April 14, 2022
Posted
August 4, 2022
Terminated
February 20, 2025
Address
668b Phillips Rd, Ofc 1, Victor, NY, 14564-9741

Description

Turbett Surgical Container, TS1500

Reason

Devices were subject to unapproved rework processes.

Action

The firm distributed recall notification to hospitals via email on 04/14/2022. The recall notification identifies the reason for recall of specific TS1500 units to be that the repair process employed could restrict the customer's ability to clean and visually check the device for cleanliness. This can be identified by the customer by evidence of bioburden or particulate buildup in the seams of the device. Inadequate or incomplete cleaning of the unit may compromise the sterility of the load. The firm asks customers to immediately discontinue use of the device and complete the provided Return Response Form to [email protected]. Turbett will contact customers to coordinate the replacement of the unit and correction of the recalled device. Customers with any questions can call 585-755-0133x106 from Monday through Friday, from 8:00 AM to 4:30 PM.

Distribution

US Nationwide Nationwide: CO, FL, NC and PA.

Quantity

7 units