FDA Recall Open, Classified

Garbin EVO, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients. (sold outside US)

Recall: Z-1500-2024 · Initiated March 19, 2024

Recall

Recall Number
Z-1500-2024
Event Number
94263
Firm
Philips Respironics, Inc.
FEI Number
2518422
Product Code
CBK
Status
Open, Classified
Root Cause
Device Design
Initiated
March 19, 2024
Posted
April 17, 2024
Address
1001 Murry Ridge Ln, Murrysville, PA, 15668-8517

Description

Garbin EVO, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients. (sold outside US)

Reason

Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.

Action

Consignees were mailed an URGENT MEDICAL DEVICE CORRECTION notice, dated 3/7/24. Consignees are asked to identify any affected product in their inventory and follow the provided instructions in order to continue using devices safely in CPAP or PSV mode. Devices that alarm are to be immediately plugged into a power source if a Loss of Power Alarm occurs. If no source of power is available, remove and replace the detachable battery. Devices running on software other than 1.05.06.00 can be updated to the latest software version, which does not experience this failure mode. The provided recall notice should be shared with all members of consignee organizations that utilize the recalled devices. If product was further distributed the provided notice should be forwarded. Consignees with any questions can contact Philips Respironics Customer Service at 1-800-345-6443 for homecare customers or 1-800-722-9377 for hospital customers.

Distribution

Domestic: Nationwide Distribution; Foreign: AE, AR, AT, AU, BD, BE, BG, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PR, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, US, UY, VN, YT, ZA, ZW.

Quantity

1,233 units