FDA Recall Terminated

InCore Lapidus Disposable Kit 28mm Left, Part Number: IC-LAP-28L - Product Usage: Intended as a three-part construct intended for internal fixation for First Metatarsocuneiform arthrodesis.

Recall: Z-1500-2021 · Initiated March 26, 2021

Recall

Recall Number
Z-1500-2021
Event Number
87665
Firm
Nextremity Solutions
FEI Number
3009540749
Product Code
HWC
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
March 26, 2021
Terminated
November 10, 2021
Address
6210 N Buffalo St, Warsaw, IN, 46580

Description

InCore Lapidus Disposable Kit 28mm Left, Part Number: IC-LAP-28L - Product Usage: Intended as a three-part construct intended for internal fixation for First Metatarsocuneiform arthrodesis.

Reason

Product could potentially contain an undersized inner diameter of the 40mm drill bushing which result in not allowing the drill bit to pass through the drill bushing, causing inability of surgeon to complete surgical procedure as intended

Action

Nextremity Solutions issued Urgent Medical Device Recall Letter to Distributors and Risk Managers on March 26,2021 stating reason for recall, health risk and action to take: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 - Medical Device Recall Return Response and send to Corporate [email protected]. This form will be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit facility's documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574- 371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected].

Distribution

AL, AK, CA, IA,IL, IN, KY ,MD. MI, NJ, NY, OR, PA, SC, TX, VA

Quantity

111 units