FDA Recall Terminated

Maquet Getinge-BEQ-T 8803#Circ Circuit Small Patient Material: 701048539

Recall: Z-1500-2019 · Initiated April 18, 2019

Recall

Recall Number
Z-1500-2019
Event Number
82703
Firm
Maquet Cardiovascular, LLC
FEI Number
2242352
Product Code
DWF
Status
Terminated
Root Cause
Device Design
Initiated
April 18, 2019
Posted
May 31, 2019
Terminated
July 28, 2025
Address
45 Barbour Pond Dr, Wayne, NJ, 07470-2094

Description

Maquet Getinge-BEQ-T 8803#Circ Circuit Small Patient Material: 701048539

Reason

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Action

Maquet Getinge issued recall notification letter on April 18, 2019 via FedEx 2-day delivery identifies issue, health risk and action to take: quarantine and return to Maquet/Getinge recalled product. Complete response form and return to acknowledge this recall. Questions contact Getinge Customer Service at 1 888 627 8383 (press option 2, then option 2) between the hours of 8 AM and 6 PM Eastern Standard Time. On 3/26/20 firm expanded recall to include an additional lot for one consignee,product: "BEQ TOP 25003 3/8 3/8 up to 7 lpm", item number 701063255R02, lot 3000087051.

Distribution

Nationwide