FDA Recall Open, Classified

Merlin 2 PCS MER37000 programmer Model MER3400 software

Recall: Z-1496-2022 · Initiated June 17, 2022

Recall

Recall Number
Z-1496-2022
Event Number
90530
Firm
St. Jude Medical, Cardiac Rhythm Management Division
FEI Number
2017865
Product Code
DXY
Status
Open, Classified
Root Cause
Software design
Initiated
June 17, 2022
Address
15900 Valley View Ct, Sylmar, CA, 91342-3577

Description

Merlin 2 PCS MER37000 programmer Model MER3400 software

Reason

The Merlin" 2 PCS model MER3700 is a portable, dedicated programming system designed to interrogate, program, display data from, and test Abbott Medical implantable medical devices during implant and follow-up.

Action

On June 16, 2022 Abbott issued a "Medical Device Correction" via: XXXX to affected consignees. The customer notification ask consignees to take the following action: 1. Prophylactic device replacement is not recommended, as device functionality, actual longevity, and ERI indicator are not impacted (device functionality remains normal and within specifications). 2. Routine follow-up should remain as per local standard of care and clinical protocol, and ERI should continue to serve as an indicator of the need for device replacement scheduling. 3. Please direct any questions about device longevity to Abbott Technical Support at 1-800-722-3774 (U.S.). 4.Upon programmer software / remote monitoring software update, the improved longevity estimate will be displayed at the patient s next interrogation. Please note that until programmers are updated, a difference in longevity estimates between programmers and remote monitoring (Merlin.net) may be observed.

Distribution

US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VI VT WA WI WV WY , Guam OUS: ALBANIA ALGERIA ANGOLA Argentina ARMENIA Australia AUSTRIA AZERBAIJAN BAHRAIN Bangladesh BELGIUM BENIN BOLIVIA BOSNIA-HERZ. BRAZIL BULGARIA Burnei Cambodia CAMEROON Canada Chile China Colombia COTE D'IVOIRE CUBA CYPRUS CZECH REPUBLIC DENMARK DOMINICAN REP. ECUADOR EGYPT EL SALVADOR ESTONIA ETHIOPIA FAROE ISLANDS FINLAND FRANCE GEORGIA GERMANY GREECE GREENLAND GUATEMALA Hong Kong HUNGARY ICELAND India Indonesia IRELAND ISRAEL ITALY JORDAN KAZAKHSTAN KOSOVO KUWAIT KYRGYZSTAN LATVIA LEBANON LIBYA LIECHTENSTEIN LITHUANIA LUXEMBOURG MACEDONIA Malaysia MALTA MAURITIUS Mexico MONGOLIA MONTENEGRO MOROCCO Myanmar Nepal NETHERLANDS New Zealand NORWAY OMAN Pakistan PALESTINE PANAMA PARAGUAY PERU Philippines POLAND PORTUGAL QATAR ROMANIA RUSSIAN FED. SAUDI ARABIA SENEGAL Singapore SLOVAKIA SLOVENIA SOUTH AFRICA South Korea SPAIN Srilanka SUDAN SWEDEN SWITZERLAND Taiwan Thailand TRINIDAD TUNISIA TURKEY UKRAINE United Arab Emirates UNITED KINGDOM UZBEKISTAN VENEZUELA Vietnam VIRGIN ISLAND YEMEN ZAMBIA

Quantity

622 programs