FDA Recall Open, Classified

Brand Name: ID Circuit Product Name: ID Circuit Flex 220 Model/Catalog Number: MP01347 Software Version: Not applicable Product Description: Breathing circuit for conveying breathing gases between an anesthesia machine or ventilator and adult Component: Not applicable

Recall: Z-1494-2025 · Initiated March 12, 2025

Recall

Recall Number
Z-1494-2025
Event Number
96342
Firm
Draeger, Inc.
FEI Number
2517967
Product Code
BZO
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
March 12, 2025
Posted
April 3, 2025
Address
3135 Quarry Rd, Telford, PA, 18969-1042

Description

Brand Name: ID Circuit Product Name: ID Circuit Flex 220 Model/Catalog Number: MP01347 Software Version: Not applicable Product Description: Breathing circuit for conveying breathing gases between an anesthesia machine or ventilator and adult Component: Not applicable

Reason

The potential for cracks forming in the breathing circuit hose.

Action

On March 12, 2025 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions to be Taken When using these breathing circuits: 1. Use holders with a larger contact surface (see Figure 2 above). 2. Perform leak testing after stretching the flexible hose to the required length for the application. Please observe the products Instructions for Use (IFU) According to our records, you have received at least one of the listed breathing circuits. Please inform all potential users in your facility. If you have forwarded affected products to third parties, please also forward this Urgent Medical Device Recall notification. Please also complete the enclosed Medical Device Recall Return Response Acknowledgment and Receipt Form to confirm to us that you have received this information. Adverse events or quality problems experienced with the use of this product may be reported to FDA s MedWatch Adverse Event Reporting program either online or by phone at 1-800-FDA-1088. We apologize for any inconveniences resulting from this action. If you have any questions regarding this letter, please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 267-664-1131 or via email at [email protected]. The US Food and Drug Administration has been advised of this action. ****Update 07/08/2025****On June 25, 2025 URGENT MEDICAL DEVICE RECALL - UPDATE letters were sent to consignees to clarify and provide complete UDI information for the device to include both primary and secondary UDI numbers as errors were identified in the initial letter.

Distribution

Domestic: Nationwide Distribution. International - following countries: Austria, Belgium, Bulgaria, China, Croatia, Czech Republic, France, Germany, Ghana, Greece, Guatemala, Hungary, India, Italy, Japan, Panama, Philippines, Portugal, Singapore, South Africa, Spain, Sweden, United Kingdom

Quantity

3325 units