EXALT Controller - Product Usage: intended for use with a Boston Scientific endoscope for endoscopic diagnosis, treatment, and video observation. UPN: M00542430
Recall
- Recall Number
- Z-1492-2021
- Event Number
- 87593
- Firm
- Boston Scientific Corporation
- FEI Number
- 3005099803
- Product Code
- FDT
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- March 16, 2021
- Terminated
- May 2, 2022
- Address
- 100 Boston Scientific Way, Marlborough, MA, 01752-1234
Description
EXALT Controller - Product Usage: intended for use with a Boston Scientific endoscope for endoscopic diagnosis, treatment, and video observation. UPN: M00542430
May have been programmed with a less than optimal brightness setting resulting in a darker than desired image
Boston Scientific issued Medical Device Removal - Immediate Action Required letter on 3/16/21. Letter states reason for recall, health risk and action to take: While no adverse health consequences are anticipated and an impacted Controller can continue to be used, replacement of the Controller is advised for optimal image quality. A Boston Scientific representative will be in contact with you directly to schedule installation of a replacement Controller at your facility. If you notice a darker than desired image, manually adjust the brightness setting on the front of the EXALT" Controller to the High setting (all five bars on the Brightness Indicator illuminated) A Boston Scientific representative will be in contact with you directly to schedule the installation of a replacement Controller at your facility. Your current Controller will be removed at that time. Please read carefully through the enclosed instructions. We ask that you please confirm receipt of this notification by completing the enclosed Acknowledgement Form and sending it to [email protected] or fax to 763-415-7708.
US Nationwide distribution in the states of CA, FL, NJ.
3 units