FDA Recall Terminated

EXALT Controller - Product Usage: intended for use with a Boston Scientific endoscope for endoscopic diagnosis, treatment, and video observation. UPN: M00542430

Recall: Z-1492-2021 · Initiated March 16, 2021

Recall

Recall Number
Z-1492-2021
Event Number
87593
Firm
Boston Scientific Corporation
FEI Number
3005099803
Product Code
FDT
Status
Terminated
Root Cause
Process control
Initiated
March 16, 2021
Terminated
May 2, 2022
Address
100 Boston Scientific Way, Marlborough, MA, 01752-1234

Description

EXALT Controller - Product Usage: intended for use with a Boston Scientific endoscope for endoscopic diagnosis, treatment, and video observation. UPN: M00542430

Reason

May have been programmed with a less than optimal brightness setting resulting in a darker than desired image

Action

Boston Scientific issued Medical Device Removal - Immediate Action Required letter on 3/16/21. Letter states reason for recall, health risk and action to take: While no adverse health consequences are anticipated and an impacted Controller can continue to be used, replacement of the Controller is advised for optimal image quality. A Boston Scientific representative will be in contact with you directly to schedule installation of a replacement Controller at your facility. If you notice a darker than desired image, manually adjust the brightness setting on the front of the EXALT" Controller to the High setting (all five bars on the Brightness Indicator illuminated) A Boston Scientific representative will be in contact with you directly to schedule the installation of a replacement Controller at your facility. Your current Controller will be removed at that time. Please read carefully through the enclosed instructions. We ask that you please confirm receipt of this notification by completing the enclosed Acknowledgement Form and sending it to [email protected] or fax to 763-415-7708.

Distribution

US Nationwide distribution in the states of CA, FL, NJ.

Quantity

3 units