FDA Recall Terminated

TCC-EZ, Total Contact Cast System, Model No. TCC2-4051

Recall: Z-1487-2018 · Initiated August 18, 2017

Recall

Recall Number
Z-1487-2018
Event Number
79636
Firm
Integra LifeSciences Corp.
FEI Number
3003418325
Product Code
ITG
Status
Terminated
Root Cause
Process control
Initiated
August 18, 2017
Terminated
July 19, 2018
Address
311 Enterprise Dr, Plainsboro, NJ, 08536-3344

Description

TCC-EZ, Total Contact Cast System, Model No. TCC2-4051

Reason

A quarantined lot of the devices was released for sale. Devices from this lot may experience increased exothermic reaction causing higher levels of heat output which exceed the product specifications.

Action

On 8/18/17, the recalling firm's sales reps notified affected consignees of the recall and distributed customer letters. Customers were asked to examine their inventory for product from the affected lot, warned not to use the product, complete the included "Acknowledgement and Return Form", and return the form by email or fax. Customers were provided with a Customer Service number: 1-800-654-2873.

Distribution

US nationwide distribution.

Quantity

250